All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.
Abstract
Simultaneous determination of valsartan and hydrochlorothiazide in tablets by RP-HPLC
Author(s): DF Tian, XL Tian, T Tian, ZY Wang, FK Mo
Department of Pharmacy, School of Pharmacy, Shenyang Pharmaceutical University, 110016 Shenyang, China
Correspondence Address:
F K Mo Department of Pharmacy, School of Pharmacy, Shenyang Pharmaceutical University, 110016 Shenyang China E-mail: fengkuimo@yahoo.com.cn
Department of Pharmacy, School of Pharmacy, Shenyang Pharmaceutical University, 110016 Shenyang, China
Correspondence Address:
F K Mo Department of Pharmacy, School of Pharmacy, Shenyang Pharmaceutical University, 110016 Shenyang China E-mail: fengkuimo@yahoo.com.cn
A simple, reproducible and efficient reverse phase high performance liquid chromatographic method was developed for simultaneous determination of valsartan and hydrochlorothiazide in tablets. A column having 200 × 4.6 mm i.d. in isocratic mode with mobile phase containing methanol:acetonitrile:water:isopropylalcohol (22:18:68:2; adjusted to pH 8.0 using triethylamine; v/v) was used. The flow rate was 1.0 ml/min and effluent was monitored at 270 nm. The retention time (min) and linearity range (µg/ml) for valsartan and hydrochlorothiazide were (3.42, 8.43) and (5-150, 78-234), respectively. The developed method was found to be accurate, precise and selective for simultaneous determination of valsartan and hydrochlorothiazide in tablets.
Full-Text | PDF