All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.
Abstract
High performance liquid chromatographic determination of sodium chromoglycate.
Author(s): AK Shakya, J Karajgi, N Talwar, AK Singhai, NK Jain
Dept. of Pharmaceutical Sciences, Dr. Harisingh Gour Vishwavidyalaya, SAGAR (M.P.) 470 003,
Correspondence Address:
A K Shakya Dept. of Pharmaceutical Sciences, Dr. Harisingh Gour Vishwavidyalaya, SAGAR (M.P.) 470 003
Dept. of Pharmaceutical Sciences, Dr. Harisingh Gour Vishwavidyalaya, SAGAR (M.P.) 470 003,
Correspondence Address:
A K Shakya Dept. of Pharmaceutical Sciences, Dr. Harisingh Gour Vishwavidyalaya, SAGAR (M.P.) 470 003
A reverse-phase HPLC assay procedure for the determination of sodium chromoglycate is described. The compound was chromatographed on a C-8 (Lichrosorb RP-8) column using a mobile phase consisting of methanol : 0.01 M K2HPO4 (3:7). Metronizadole was used as internal standard. The chromatographic system exhibited good precision (cv < 2 percent). The recovery of drug from formulation was quantitative (99-102).